Secure transport of cells without electric mains and CO2 connection
Factory-aligned parameters: 37 °C and 5 % CO2
Different temperature or CO2 concentration possible (to be advised before order)
Operation at 12 V car voltage possible
Use of different heating blocks
Easy handling and cleaning
Compact Design
Parameters related to an ambient temperature of 23 °C during transport without connection to electric mains and CO2 bottle:
Time
Temp.
CO2
0 min
37 °C
5 %
10 min
36.6 °C
4.9 %
30 min
35.5 °C
4.8 %
60 min
34.0 °C
4.7 %
Technical Specifications
Power supply
115 V or 240 V AC, 50/60 Hz 12 V DC with car voltage via adapter cable
Dimensions
Outside: 385 × 245 × 305 mm (l × w × h) Inside: approx. 312 × 167 × 140 mm (l × w × h)
Weight
Approx. 4.7 kg
Temperature
Available between 32 °C and 39 °C (to be advised before order)
Accuracy of regulation
± 0.3 °C (measured with heating block REF 10264)
Heating up time
Approx. 2 h
CO2 concentration
5 % (different concentrations possible) Heating block inside necessary Connection to CO2
supply with pre-pressure of 0.8 bar necessary CO2 accuracy ± 0.3 % CO2 (when supplied